The MIC 10 offers you various possibilities of instru-
ment configuration using a special mode. You can
compile the function range for the instrument if you
wish. You can
•
select the language for the report printout
•
select the type of report to be printed
•
inhibit individual scales
•
select whether conversion is to be made according
to DIN 50150 or ASTM E 140
•
inhibit calibration changes
•
inhibit setting the dwell time
•
inhibit adjustments to the alarm thresholds
•
adjust the backlight period
•
inhibit the function Data Logger or Memory Card.
Basic procedure
–
Select the
Configuration
level, by simultaneously
pressing
+
and
M
when the instrument is turned
on.
–
Select the different functions after each other in the
Configuration level with
M
.
–
Carry out the function adjustments with
^V
.
–
You can reset each function back to the last setting
which was stored by pressing
+
.
You can exit any function for the Configuration level
and return to the
MEASUREMENT
level by pressing
-
.
The instrument automatically switches off after 3
minutes if no key is pressed.
When you are in the configuration level the minus sym-
bol flashes in the display.
Configuration
6-2
Issue 08, 08/01
Krautkrämer MIC 10
The MIC 10 offers you various possibilities of instru-
ment configuration using a special mode. You can
compile the function range for the instrument if you
wish. You can
•
select the language for the report printout
•
select the type of report to be printed
•
inhibit individual scales
•
select whether conversion is to be made according
to DIN 50150 or ASTM E 140
•
inhibit calibration changes
•
inhibit setting the dwell time
•
inhibit adjustments to the alarm thresholds
•
adjust the backlight period
•
inhibit the function Data Logger or Memory Card.
Basic procedure
–
Select the
Configuration
level, by simultaneously
pressing
+
and
M
when the instrument is turned
on.
–
Select the different functions after each other in the
Configuration level with
M
.
–
Carry out the function adjustments with
^V
.
–
You can reset each function back to the last setting
which was stored by pressing
+
.
You can exit any function for the Configuration level
and return to the
MEASUREMENT
level by pressing
-
.
The instrument automatically switches off after 3
minutes if no key is pressed.
When you are in the configuration level the minus sym-
bol flashes in the display.
Configuration
6-2
Issue 08, 08/01
Krautkrämer MIC 10
Summary of Contents for Krautkramer MIC 10
Page 6: ...0 6 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 7: ...1 Introduction Krautkrämer MIC 10 Issue 08 08 01 1 1 1 Introduction ...
Page 16: ...1 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 29: ...4 Basics of operation Krautkrämer MIC 10 Issue 08 08 01 4 1 4 Basics of operation ...
Page 37: ...5 Operation Krautkrämer MIC 10 Issue 08 08 01 5 1 5 Operation ...
Page 69: ......
Page 70: ......
Page 71: ...6 Configuration Krautkrämer MIC 10 Issue 08 08 01 6 1 6 Configuration ...
Page 88: ...7 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 89: ...8 Care and maintenance Krautkrämer MIC 10 Issue 08 08 01 8 1 8 Care and maintenance ...
Page 97: ...10 Specifications Krautkrämer MIC 10 Issue 08 08 01 10 1 10 Specifications ...
Page 100: ...10 4 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 123: ...12 Appendix Krautkrämer MIC 10 Issue 08 08 01 12 1 12 Appendix ...
Page 133: ...13 Changes Krautkrämer MIC 10 Issue 08 08 01 13 1 13 Changes ...
Page 135: ...14 Index Krautkrämer MIC 10 Issue 08 08 01 14 1 14 Index ...