Measuring with defined measurement
times (dwell time)
As opposed to short time measurements, you can
carry out measurements whose dwell time you can
determine yourself. In order to measure with a steady
accuracy you require a cylindrical probe attachment
as well as a test support.
You can set a dwell time between 1-99 seconds.
Setting the dwell time
–
Press
M
until the following indication appears in the
display:
–
Set the required dwell time with
^8
, for example:
–
Press
-
if you wish to return to the measurement
mode or simply carry out a measurement.
Carrying out a measurement
–
Screw on the cylindrical attachment tightly down to
screw stop on the probe case.
–
Clamp the probe into the support (DH 191 or
MIC 222). By doing this you will avoid measurement
inaccuracies.
–
Select the required hardness scale.
–
If necessary set the calibration value and the alarm
threshold.
–
Carry out the measurement.
–
Make certain that the tip of the probe attachment is
pressed on the material to be measured during the
complete measurement time.
Operation
Hardness measurement
5-18
Issue 08, 08/01
Krautkrämer MIC 10
Measuring with defined measurement
times (dwell time)
As opposed to short time measurements, you can
carry out measurements whose dwell time you can
determine yourself. In order to measure with a steady
accuracy you require a cylindrical probe attachment
as well as a test support.
You can set a dwell time between 1-99 seconds.
Setting the dwell time
–
Press
M
until the following indication appears in the
display:
–
Set the required dwell time with
^8
, for example:
–
Press
-
if you wish to return to the measurement
mode or simply carry out a measurement.
Carrying out a measurement
–
Screw on the cylindrical attachment tightly down to
screw stop on the probe case.
–
Clamp the probe into the support (DH 191 or
MIC 222). By doing this you will avoid measurement
inaccuracies.
–
Select the required hardness scale.
–
If necessary set the calibration value and the alarm
threshold.
–
Carry out the measurement.
–
Make certain that the tip of the probe attachment is
pressed on the material to be measured during the
complete measurement time.
Operation
Hardness measurement
5-18
Issue 08, 08/01
Krautkrämer MIC 10
Summary of Contents for Krautkramer MIC 10
Page 6: ...0 6 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 7: ...1 Introduction Krautkrämer MIC 10 Issue 08 08 01 1 1 1 Introduction ...
Page 16: ...1 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 29: ...4 Basics of operation Krautkrämer MIC 10 Issue 08 08 01 4 1 4 Basics of operation ...
Page 37: ...5 Operation Krautkrämer MIC 10 Issue 08 08 01 5 1 5 Operation ...
Page 69: ......
Page 70: ......
Page 71: ...6 Configuration Krautkrämer MIC 10 Issue 08 08 01 6 1 6 Configuration ...
Page 88: ...7 10 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 89: ...8 Care and maintenance Krautkrämer MIC 10 Issue 08 08 01 8 1 8 Care and maintenance ...
Page 97: ...10 Specifications Krautkrämer MIC 10 Issue 08 08 01 10 1 10 Specifications ...
Page 100: ...10 4 Issue 08 08 01 Krautkrämer MIC 10 ...
Page 123: ...12 Appendix Krautkrämer MIC 10 Issue 08 08 01 12 1 12 Appendix ...
Page 133: ...13 Changes Krautkrämer MIC 10 Issue 08 08 01 13 1 13 Changes ...
Page 135: ...14 Index Krautkrämer MIC 10 Issue 08 08 01 14 1 14 Index ...