Setup Menu Options
|
53
MicroFlo Select Microplate Dispenser
Setup Menu Options
Dispenser Action Menu> SETUP
SETUP OPTONS:
CAL CASS BioStack
Top line
Second line
Softkeys
The Setup menu holds configuration and other critical details:
CAL
: short for calibrate, is an aid to calibrating the MicroFlo on our factory
floor. It is not needed when calibrating cassettes gravimetrically or when
using the Cassette Calibration Kit.
CASS
: short for cassette, tells the MicroFlo which cassette type is installed.
This is critically important for expected performance. The size of the tubing
determines the amount of fluid dispensed per well.
BioStack
: offers controls for using the BioStack to process plates with the
MicroFlo. See
on page 55.
Pattern
: is only applicable when processing 384- and 1536-well plates. It
lets you select the fill pattern:
Column
or
Row
. Select Row for faster
throughput.