Using the BioStack
|
55
MicroFlo Select Microplate Dispenser
Using the BioStack
Dispenser Action Menu> BIOSTACK
If you have purchased BioTek’s BioStack™ Microplate Stacker to operate with the
MicroFlo, follow the instructions in your
BioStack Operator’s Manual
to set it up to
interact with the dispenser.
After installing the BioStack to operate with the MicroFlo, use the MicroFlo’s menu to:
Configure
: tell the MicroFlo to use the BioStack.
Align
: align the BioStack’s gripper with the MicroFlo’s plate carrier.
Verify
: test the setup.
Set the MicroFlo to use the BioStack:
After installing the BioStack, tell the MicroFlo to use it to process plates:
1.
At the main menu, press
twice to get to the
Dispenser Action Menu
2.
Select
SETUP
3.
Select
BioStack
4.
Select
Conf
0.
BIOSTACK UTILITIES:
CONF ALIGN VERIFY
Top line
Second line
Softkeys
5.
At the Operating Mode menu, select
BioStack
.
Repeat these steps, except select
Manual
, to run the MicroFlo without the BioStack.
Align the BioStack with the MicroFlo:
The BioStack’s installation instructions include an alignment step. Use the MicroFlo’s
menu to access the BioStack Alignment Utility:
1.
At the main menu, press
twice to get to the
Dispenser Action Menu
2.
Select
SETUP
3.
Select
BioStack
4.
Select
Align
0.
Follow instruction provided in the BioStack Operator’s Manual.
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...