50 | Operation
BioTek Instruments, Inc.
Define the Cassette
Dispenser Action Menu> SETUP> CASS
It is important to correctly identify the cassette type for accurate dispensing. The current
cassette setting is shown in the main menu.
(384) CASS(10) VOL:0010µL
Prime Purge uPlate
CASS(#) in the top line of the LED displays the current Cassette setting. Likewise, the
cassette is prominently labeled. Make sure the MicroFlo’s setting matches the currently
used cassette.
The MicroFlo displays a warning message when an out-of-range volume is defined for the
current cassette. For example, if you request a 150 µL dispense with a 1 µL cassette, a
message will require you to confirm your intention or cancel the run.
To define the cassette type:
1.
At the MicroFlo’s main menu, select
twice to get to the
Dispenser Action
Menu
.
2.
Select
SETUP
.
3.
Select
CASS
.
4.
Select the
1 µL
,
5 µL
, or
10 µL
cassette.
0.
This action returns you to the Setup Options. The cassette type is changed to
your selection. Confirm this at the main menu: press
Main Menu
twice.
Removing the Tubing Cassette
Review the Tubing Cassette Diagram on page 14.
1.
Lift the
Tube Organizer
out of the fluid vessel.
2.
Move the
Pump Cover
away from the pump to its off position.
3.
Release the stainless steel plate holding the
Tube Tensioner
: release the catch on
the top front of the pump that holds it in place. It will hang loosely.
4.
Lift the
Center Holder
out of its slot on the right side of the pump.
5.
Hold the Center Holder in one hand while you slide the
Tip Holder
out of the
dispense arm.
0.
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...