80 | Chapter 6: Instrument Qualification
© 2008, BioTek Instruments, Inc.
MicroFlo Select Dispense Precision & Accuracy Test Worksheet
1 µL Cassette Test
1 µL Dispense Precision Test
Standard Deviation (SD):
Mean Absorbance (sum of all wells
96)
% CV (SD
Mean x 100)
%
% CV must be < 11.0%
Pass
Fail
1 µL Dispense Accuracy Test
Total Dispense (Actual) Weight:
grams
Expected Weight:
(mL/well x number of wells dispensed)
grams
% Accuracy Error:
(Actual Weight – Expected Weight)
Expected Weight x 100
%
% Accuracy Error must be < 10.0%
Pass
Fail
Visual verification that no well varies
considerably from the others
Pass
Fail
Cassette Serial Number:
Tests Performed By:
Date:
Reviewed/Approved By:
Date:
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...