96 | Dispensing to Test Tubes
BioTek Instruments, Inc.
Install the test tube rack
You must swap the microplate carrier for the test tube rack to dispense to test tubes
with the MicroFlo:
3
Turn off the MicroFlo.
4
Optionally, disconnect the drain tube from the prime trough.
5
Remove the prime trough by sliding it towards you and away from the
dispenser.
6
Move the plate carrier all the way to the left and lift it up and off its rails. Lift
the carrier’s left side first.
7
Place the test tube rack on the rails with the engraved FRONT label oriented to
the front of the instrument. Notches in the base of the test tube rack secure it in
place. Make sure the rack is level.
8
Reinstall the priming trough and drain tube.
0
Note: Only four tubes are used to dispense to the test tube rack.
Tubes 1, 3, 5, 7 must be removed from the fluid vessel before
beginning a dispense.