Performing the Precision and Accuracy Tests
|
77
MicroFlo Select Microplate Dispenser
Performing the Precision and Accuracy Tests
Prerequisite:
Prepare the test solutions as described on page 76.
Make a copy of the applicable worksheet beginning on page 80.
Procedure:
1.
Install the cassette to be tested.
2.
Turn on the MicroFlo Select and make sure the cassette type setting is correct.
3.
Turn on the balance.
4.
Fill a beaker or other vessel with the test solution.
5.
Define a
Protocol
and save it for reuse. Set the parameters based on the desired
test volume:
Set the
Dispense Volume
to match the Test:
Set the
Flow Rate
to:
1 µL cassette = Medium
5 µL cassette = High
10 µL cassette = High
Retain the default settings for Dispense Height, Plate Clear Height and
Horizontal Position.
Define a
Prime before start
: set the volume to 10 µL.
Set the
Number of Pre-dispenses
to 2.
1536 Test
only: : Change the
plate map
to dispense to every other column.
It will be easier to observe the dispense results than dispensing to the whole
plate.
6.
Place a clean/new microplate on the balance and tare the balance.
Test
Volume
1 µL
1 µL
5 µL
5 µL
10 µL
10 µL
50 µL
50 µL
100 µL
100 µL
1536
6 µL
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...