Unclogging the Dispense Tips
|
63
MicroFlo Select Microplate Dispenser
Unclogging the Dispense Tips
The small diameter of the dispense tips makes them susceptible to clogging. You may be
able to visually identify a clogged tip or inaccurate dispense performance may signal a
problem. Good work habits can prevent clogging or reduce its occurrence:
When using 1 µL cassettes, filter fluids to 50 microns before dispensing.
Thoroughly flush the tubing after/in-between usage, especially when using
liquids that crystallize or harden.
In case the need arises, BioTek ships a 10 cc plastic syringe with special tubing and
fitting for use unclogging tips. Installation instructions recommend storing it in the
pouch on the back of the instrument. The remedy involves removing the dispense tip
and flushing it with water. Depending on the type of clog, soaking the tip holder in hot
water with mild detergent is recommended.
This task may be easier if you use the cassette’s shipping container to hold the
unaffected cassette parts, keeping them out of your way.
Required Materials
10 cc syringe with tubing and fitting attachment shipped with dispenser
Screwdriver shipped with dispenser