36 | Essential Concepts
BioTek Instruments, Inc.
Table 2. Chemical Compatibility Ratings
Key
1 2 3 4 5 6 7 8 9 10
A
No effect
B Slight
effect
C Moderate
effect
D Severe
effect
ND No
data
316 S.Steel
Alumi
num
Brass
Polycarbon
.
Polypropy
l.
Polyst
yren
e
Ptf
e
Ryton
Silicon
e
Ultem
Chemical
(1-10 represent Materials listed in Table 1 above)
1
2
3
4
5
6
7
8
9
10
Acetic Acid, 5%
A B D A A
D A A A
A
Acetic Anhydride
A A D D B
D A A A ND
Acetonitrile
A B ND D A
D A A D
(1)
D
Ammonia 10%
A A ND D A
B A
A D
D
Benzyl Alcohol
B B ND ND A D A A A ND
Chloroform A
B B D C D
A A D
D
Detergents 1%
A B ND A A
A A A A
A
Dimethylformamide B
A ND D A
D A A A ND
DMSO (Dimethylsulfoxide)
A A ND D A
D A A C
(1)
D
Ethyl Alcohol 70%
A A A B A
A A A B
A
Ethylene Oxide
B D D C D
C A D A
ND
Formaldehyde 37%
A B A A A
ND A A C A
Hexane A
A A D B
D A A D
A
Hydrocholoric Acid 20%
D D D B B
C A D D
A
Hydrofluoric Acid 20%
D D ND D A
ND A A D ND
Hydrogen Peroxide 10%
B A D A A
A A C A
A
Isopropyl Alcohol 70%
A A A A A
A A A A
A
Methyl Alcohol 70%
A A A B A
ND A A A
A
Methylene Chloride
B C A D B
D A A D
D
Phosphoric Acid 10%
D C D A A
B A
A C
A
Propylene Glycol
B B ND B A
A A A A ND
Sodium Chlorate
B C ND A A
ND A A C ND
Sodium Hydroxide 20%
B D B A A
A A A A
A
Sodium Hypochlorite <20% C D D C A
A A C B
B
Sodium Hypochlorite 0.5% B D D C A
A A C B
A
Sulfuric Acid <10%
B B ND A A
A A A C A
Trichloroethylene B
D ND ND C D
A A D
D
Virkon 10%
A D ND A A
A ND A A ND
(1) Exposure to DMSO and Acetonitrile may cause the silicone tubing
to swell, increasing the volume of fluid dispensed. The magnitude of
this effect will vary with concentration and exposure. Re-calibration of
the cassette may be required.
Please contact BioTek with questions about compatibility of any
chemicals not described here.
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...