About Calibrating Cassettes Gravimetrically
|
107
BioTek Instruments, Inc.
Autoclaving the cassette may change (increase) the tubing's capacity. After
multiple autoclave cycles, recalibration may be required.
When working with a newly built cassette, use the numbered scale on the Tube
Tensioner’s face to evenly line up the blocks. Calibration will be easier when all
the tubes begin at the same point.