![Biotek MicroFlo select Operator'S Manual Download Page 70](http://html.mh-extra.com/html/biotek/microflo-select/microflo-select_operators-manual_2756071070.webp)
52 | Operation
BioTek Instruments, Inc.
The MicroFlo’s
Plate Map
defines which column or range of columns to dispense to. The
graphic representation of the columns makes it easy to pick and choose the plate’s
columns. Plate Maps can be changed on-the-fly for both
Quick Dispenses
and
Protocols
.
Plate Types Table
Dispense Height
Plate Type
Plate
Name
Columns x
Rows
Steps
mm
Tubes
per Well
96 Well
96
12x8
336
15.36
1
384 Well
384
24x16
336
15.36
1
1536 Well
1536
48x32
254
11.61
1
384 PCR
384P
24x16
254
11.61
1
96 Mini Tubes
96MT
12x8
1124
51.40
1
96 Deep Well
96D
12x8
964
44.07
1
384 Deep Well
384D
24x16
964
44.07
1
20 12x75 Tubes
75T
5x4
1433
65.52
1*
20 13x100 Tubes
100T
5x4
1989
90.94
1*
6 Well
6
3x2
488
22.31
4
12 Well
12
4x3
488
22.31
2*
24 Well
24
6x4
488
22.31
2
48 Well
48
8x6
488
22.31
1*
*Important
: when dispensing to test tubes, and 6-, 12- and 48-well plates some dispense
tubes must be removed from the fluid supply vessel. See
Special Plates Handling
on page
33.
Fill Pattern
When using
384-
and
1536-well
plates, the MicroFlo offers two dispense patterns: Column
or Row. For these high-density plates, the 8-tip manifold must address the plate multiple
times to fill it. Column-wise dispensing fills each column before moving to the next. Row-
wise dispensing fills the first 8 rows, then reverses direction to fill the next 8 rows, and so
on, if applicable. Row-wise processing is faster because it requires fewer changes in
direction. Once it is defined, your
Pattern
preference will apply to all runs, Quick
Dispenses and Protocols.
Using the keypad
: Set the fill pattern using the
Dispenser Action Menu>Setup
.
Using the LHC
: Select
Tools>Instrument Utilities
to set the fill pattern
Test Tube Racks
BioTek provides a test tube rack for use with the MicroFlo as optional accessory. It
supports dispensing to two common test tube sizes. Contact your BioTek representative to
purchase one.
Be sure to follow instructions for removing the plate carrier and properly replacing it with
the test tube rack in
Appendix B
.
Summary of Contents for MicroFlo select
Page 1: ...Operator s Manual Dispenser MicroFlo Select ...
Page 2: ......
Page 44: ...26 Installation BioTek Instruments Inc ...
Page 74: ...56 Operation BioTek Instruments Inc ...
Page 116: ...98 Dispensing to Test Tubes BioTek Instruments Inc ...
Page 126: ...108 Gravimetric Calibration Instructions MicroFlo Select Microplate Dispenser ...