Alarm and Alarm Status
High P-wave and T-wave, Long P-wave or T-wave with high amplitude duration can be detected
by QRS Complex. Place the leads on the ECG1 channel with the highest R-wave (compare to T-
wave and / or P-wave) to allow the monitor to properly detect low heart rate conditions in this
situation. If the monitor continues to misinterpret the P-wave or T-wave, use a pulse oximeter to
reposition the electrodes or monitor the animal's pulse rate.
Display
①
②
③
④
⑤
①
Heart rate detector: It detects heart rate and flickers simultaneously.
②
Pace maker: Pace maker signal is detected and flashes simultaneously.
③
Heart rate: Displays the heart rate per minute.
④
HR Alarm limits: Heart rate upper and lower thresholds are displayed.
⑤
PVC count number per 1 minute is display.
ECG Settings
Main menu
Sub menu
ECG
A.
ECG Parameters
A-1.
Alarm
A-2
. QRS Volume
A-3.
Display Option
A-4.
Arrhythmia
A-5.
ST/PVC
A-6.
Pace Maker