79
BM7Vet Pro
User’s Manual
Warning
Display Heart Beat Equipment Signal
Heart Beat equipment signal displays when the PACE mode is. the signal appears series
form. The signal size or form are meaningless clinically
Number Of Heart Beat
Attention to the animal with heart beat equipment. The heart beat equipment can show
heart beat even during arrhythmia continuously. Therefore, do not depend on heart beat
alarm excessively.
Warning
VENTRICULAR ARRHYTHMIAS
The arrhythmia analysis program is intended to detect ventricular arrhythmia. This program
is not designed to detect trial or supra ventricular arrhythmias. In some cases, it may not be
possible to distinguish the presence or absence of arrhythmias. Therefore, doctors should
analyze the arrhythmia information like other medical information.
SUSPENDED ANALYSIS
Certain conditions can delay the arrhythmia analysis. Detection and alarms associated with
arrhythmias do not occur when arrhythmia conditions are delayed. This message is
generated when the arrhythmia analysis is delayed:
LEADS FAULT, ALARM PAUSE, ALL ALARMS OFF, DISCHARGED.
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...