184
BM7Vet Pro
User’s Manual
Note 1)
U
т is the
A.C. mains voltage prior to application of the test level.
Note 2)
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 3)
These guidelines may not apply in all situations. Electromagnetic propagation is affec
ted by absorption and reflection from structures, objects and people.
a
Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) tele
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcas
t cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea
sured field strength in the location in which the EUT is used exceeds the applicable RF comp
liance level above, the EUT should be observed to verify normal operation. If abnormal perfor
mance is observed, additional measures may be necessary, such as re-orienting or relocating
the EUT.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V /
m.
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and
the BM7 Vet Pro system.
The BM7 Vet Pro system is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The user of the BM7 Vet Pro system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the BM7 Vet Pro system as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5
GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...