40seconds. The typical time for a zero
is 15~20seconds.
C-1.
MODULE RESET
EtCO
2
MODULE initializing.
Note
For best result, connect the CAPNOSTAT 5 CO
2
Sensor to an adapter and wait 2 minutes
before performing the Adapter Zero procedure.
Status Message
Following is a list of some of the message that may appear on the monitor when monitoring CO
2
.
The message should clear when normal operating criteria are met or a solution is found.
* SENSOR OVER TEMP
- Cause: The sensor temperature is greater than 104
° F
- Solution: Make sure sensor is not exposed to extreme heat (heat lamp, etc.)
* SENSOR FAULTY
- Cause: One of the following conditions exist: Capnostat Source Current Failure
EEPROM Checksum Faulty, Hardware Error
- Solution: Check that the sensor is properly plugged in. Reinsert or reset the sensor if
necessary.
* SENSOR WARM UP
- Cause: Sensor under temperature, temperature not stable, Source Current unstable
- Solution: This error condition is normal at startup. This error should clear when the warmup is
complete.
* CHECK SAMPLING LINE
- Cause: This error occurs whenever the pneumatic pressure is outside the expected range.
- Solution: Check that the sampling line is not occluded or kinked. Replace the sample line
* ZERO REQUIRED
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...