14
BM7Vet Pro
User’s Manual
Electromagnetic Compatibility
The monitor is designed and tested to comply with current regulatory standards (IEC
60601-1-2 and CISPR 11 Class A) on its ability to reduce electromagnetic emissions (EMI)
and to isolate EMI from external sources.
To reduce possible problems caused by electromagnetic interference, we recommend the
following:
⚫
Use only Bionet approved accessories. Otherwise, this equipment may not work
correctly.
⚫
Ensure that other products used in areas where veterinary patient monitoring
and life support is used comply to accepted emissions standards (CISPR 11, Class
A).
⚫
Try to maximize the distance between electromedical devices. High-power
equipment related to electrical simulators, electrosurgical instruments and
radiators (X-ray machines) as well as evoked potential devices may cause monitor
interference.
⚫
Strictly limit exposure and access to portable radio frequency sources (e.g.
cellular phones and radio transmitters). Be aware that portable phones may
periodically transmit even when in standby mode.
⚫
Maintain good cable management. Do not route cables over electrical equipment.
Do not intertwine cables.
⚫
Ensure all electrical maintenance is performed by qualified personnel.
Caution
Infectious devices and parts must be sanitized and cleaned before disposal.
사용
안내서
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...