16
BM7Vet Pro
User’s Manual
Caution
The Equipment should be placed far from generator, X-ray equipment, broadcasting
equipment or transmitting wires, so as to prevent electrical noises from being generated
during the operation, When these devices are near the Equipment, it can produce
inaccurate measurements. For BM7 Vet Pro both independent circuit and stable grounding
are essentially required. In the event that the same power source is shared with other
electronic equipment, it can also produce inaccurate output.
Note
BM7 Vet Pro is classified as follows:
- BM7 Vet Pro classifies as Class
II,
BF
&
CF concerning electric shock. It is not proper to
operate this Equipment around combustible anesthetic or dissolvent.
- Noise level is A class regarding IEC/EN 60601-1 and the subject of Nose is A level
concerning IEC/EN60601-1-2.
Warning
Do not touch the animal while using the defibrillator. The user may be at risk.
When using the defibrillator, be careful about safety and use only the supplied cable.
Warning
In case the Equipment does not operate as usual or damaged, do not use on animal, and
contact to the medical equipment technician of the hospital or the equipment supply
division.