181
BM7Vet Pro
User’s Manual
Voltage dips, sh
ort
Interruptions an
d
Voltage variation
s
on power suppl
y
input lines
IEC 61000-4-11
<5% U
т
(>95% dip in U
т)
for 0.5cycle
40% U
т
(60% dip in U
т)
for 5 cycle
70% U
т
(30% dip in U
т)
for 25 cycle
<5% U
т
(<95% dip in U
т)
for 5 s
<5% U
т
(>95% dip in U
т
)
for 0.5cycle
40% U
т
(60% dip in U
т)
for 5 cycle
70% U
т
(30% dip in U
т)
for 25 cycle
<5% U
т
(<95% dip in U
т
)
for 5 s
Mains power quality sh
ould be that of a typic
al commercial or hospit
al environment. If the u
ser of the BM7 Vet Pro
system requires continu
ed operation during po
wer mains interruptions,
it is recommended that
the BM7 Vet Pro system
be powered from an un
interruptible power sup
ply
or a battery
Note:
U
т is the a.c. mains voltage prior to application of the test level.