168
BM7Vet Pro
User’s Manual
SpO
2
Malfunction
-
If the connectors of the equipment are in bad condition
The extension cables
Are disconnected
The finger probe is
in bad condition?
No
No
Replace extension cables
Replace the finger probe
Repair the ECG B/D
Yes
Yes
Temperature malfunction
- If the temperature cannot be measured, check the connection with the equipment
The probe is
In bad condition?
No
Replace the probe
Repair the Temp B/D
Yes