A-10-2. Address 1
Address information 1
A-10-3. Address 2
Address information 2
A-10-4. Postal Code
Set postal Code
A-10-5. Doctor Name
Set Doctor Name
B. Biosignal Calibration
Set calibration menu
B-1. ECG & RESP
B-1-1. ECG Calibration
ECG calibration menu
10mm/mV input calibration
display
B-1-2. RESP Calibration
RESP calibration menu
1ohm 1cmm display
B-2. NIBP
B-2-1. ZERO Calibration
NIBP Zero calibration menu
Zero calibration menu at
atmospheric pressure
B-2-2.
Gain Calibration
NIBP Gain control menu
Perform 250mmHg pressure
calibration and select menu
B-2-3.
Pneumatic Pump
NIBP Pump control menu
ON/ OFF
B-2-4.
Pneumatic Valve
NIBP valve control menu
Close /Open
B-3.
IBP
B-3-1. IBP1 Calibration
Perform 100mmHg pressure
calibration and select menu
B-3-2. IBP2 Calibration
Perform 100mmHg pressure
calibration and select menu
C. Screen Calibration
Perform touch screen
calibration point input
D. Maker Services Setting
D-1. MAC Address
Editing
Enter a unique address for the
device
D-2. MGAS Services
E. SW Upgrade
Software Upgrade Menu
F. SW LICENSE
SW License menu