180
BM7Vet Pro
User’s Manual
M
anufacturer’s declaration
- electromagnetic immunity
The BM7 Vet Pro system is intended for use in the electromagnetic environment specified below
.
The customer or the user of the BM7 Vet Pro system should assure that it is used in such an
environment
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
Environment -guidance
Electrostatic disc
harge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood,
concrete or ceramic tile
. If floors are covered
with synthetic material,
the relative humidity sh
ould be at least 30 %
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply line
s 1kV for input/output lin
es
2kV for power supply lin
es
1kV for input/output line
s
Mains power quality sh
ould be that of a typic
al commercial or hospit
al environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality sh
ould be that of a typic
al commercial or hospit
al environment.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency magne
tic fields should be at l
evels characteristic of a
typical location in a typ
ical commercial or hosp
ital environment.
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...