94
BM7Vet Pro
User’s Manual
NIBP Settings
A.
NIBP menu
Menu
Description
Available Settings
A-1.
Alarm
NIBP Alarm setup menu
A-1-1.
Parameter Alarm Limit
SYS, MEAN, DIA Parameter alarm limit,
level, activation setup
A-1-2.
TECHNICAL ALARM
CONDITION
NIBP-OVER PRESSURE
NIBP-OVERTIME PRESSURE
NIBP-INFLATION FAILURE
NIBP-DEFLATION FAILUER
NIBP-MEASUREMENT ERROR
NIBP-PULSE TOO WEAK
NIBP-AIRLEAK,
NIBP-EXCESSIVEMOTION,
NIBP-SYSTEMFAULT
A-2.
Cuff Size
A menu to select cuff size
* When changing the cuff, be sure to
adjust the actual size of the cuff
and the size of the cuff set on the
equipment.
SMALL (1-2)
MEDIUM (3-4)
LARGE (5>)
A-3.
Inflation
A function to set the range that is
usually used by setting pressure at the
beginning because it can cause pain
for the animal when the equipment is
turned on and pressurized to the
maximum pressure range at the initial
pressurization.
Default Settings value:
SMALL: 60-140mmHg
MEDIUM: 80-
140mmHg
LARGE: 120-
250mmHg