130
BM7Vet Pro
User’s Manual
with male Luer Lock connector and the Nomoline Adapter with female Luer Lock connector.
The Nomoline Airway Adapter Set with integrated airway adapter can be used with intubated
animals. The Nomoline with a male Luer Lock type connector is compatible with any normal
configuration that uses a female Luer Lock connector. When connecting to a T-adapter, be sure to
use a PHASEIN T-adapter that samples the gas from the center of the T-adapter (see below).
For optimal water handling, always use T-adapters with the sampling point in the center of the adapter, as
shown to the left in the figure above.
The Nomoline Adapter with female Luer Lock connector connects to a standard male Luer to Luer
sample line (Nomo Extension) as well as to different kinds of third-party cannulas for oral and
nasal sampling. Combining the Nomoline Adapter with the Nomo Extension and T-adapter results
in a similar product as the Nomoline Airway Adapter Set as show below
Combining Nomoline Adapter with the Nomo Extension and T-adapter results in a similar product
If using third-party sample tubes or cannulas, make sure that the inner diameter does not exceed 1 mm
since this will increase the ISA’s total system response time.
Note
Using sample tubes or cannulas with larger inner diameter than 1 mm will increase
ISA’s
total system response time.
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...