105
BM7Vet Pro
User’s Manual
1)
Close the transducer stopcock on the animal
’s side.
2)
Open the venting stopcock on the air side.
3)
Press the knob switch on the monitor panel.
4)
Draw a line with the current input data in IBP area of WAVE WINDOW according to the Wave
Base Line. And accord the wave line with the data.
5)
Set the data as ‘0’ on the parameter screen.
6)
Check if Zero reference is carried out. (Check the pressure parameter on the message
window.)
7)
Close the venting stopcock on the air side.
8)
Open the transducer stopcock on the animal side. The pressure value should be displayed on
the pressure parameter screen in a few seconds.