43
BM7Vet Pro
User’s Manual
and appropriate action must be taken.
Subsequent changes to the network example:
Network configuration change
Removing a device from the network
Adding new devices to the network
Upgrading or updating devices connected to the network
Warning
Recommendations for wireless connections
⚫
BM7 Vet Pro
has a change in the number of equipment connections depending on
wireless AP (Access Point) performance.
⚫
When using a general AP, it is recommended to connect 8 units to the same network.
⚫
Due to the nature of wireless, connectivity may not be good depending on the
environment
.
⚫
If you want to use B2B function, Wired LAN is recommended.
IT Network connection
No one other than service personnel can connect this device to your network. Please consult with
the hospital IT staff in advance. Please refer to the following documents to proceed with the
installation.
⚫
Documents attached to this unit
⚫
Network Interface Manual
⚫
BM Vet Central (BT-Link, B-Link) user documentation
We recommend that you follow IEC 80001-1 (Hazard Management of IT Networks Connected with
Medical Devices).
LAN Network
LAN networks are usually configured through a star topology. Individual devices can be combined
into groups via a layer-n-switch. Other data traffic is separated by separate VLAN networks.
Configure the device's network settings according to this user manual and network specifications.
LAN connection specifications are described in the following standard specifications.
⚫
Wired Network: IEEE 802.3
⚫
Wireless network: IEEE 802.11 (a, b, g, n)
If the device is to be used as a layer-2-switch or layer-3-switch, the port setting must be
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...