Trouble Shooting and Solutions
When the measurement density error is high during operation
①
Check the connection of sampling line and water trap and reconnect.
②
Check the connection between the water trap and the water trap holder and reconnect them.
③
Perform manual calibration in an unpolluted environment.
④
If you close the gas input part of the water trap by hand and operate it for a few seconds,
“Occlusion”
will appear on the host device. At this time, the motor operates strongly and absorbs
gas from the gas input part. If the above operation does not appear, please contact our head
office for inspection.
Regular inspection
As with all medical equipment, we encourage you to have your device tested regularly once a
year for BMGA.