67
BM7Vet Pro
User’s Manual
ESU, RF interference may affect the monitor operation.
✓
Locate the monitor as far as possible from the ESU. Locate them on opposite sides of
the operating table, if possible.
✓
Connect the monitor and ESU to different AC outlets located as far as possible from
each other.
✓
When using this monitor with an electrosurgical unit, its return plate and the electrodes
for monitoring must be firmly attached to the animal. If the return plate is not attached
correctly, it may burn the animal
’
s skin where the electrodes are attached.
During surgery
Use the appropriate orange electrode ECG safety cable, or lead cable with a red connector,
for measuring ECG in the operating room. These cables have extra circuitry to protect the
animal from burns during cautery, and they decrease electrical interference. This also
reduces the hazard of burns in case of a defective neutral electrode at the HF device. These
cables cannot be used for measuring respiration.
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...