189
BM7Vet Pro
User’s Manual
Indicator
Categorized alarms (3 priority levels), Visual alarm lamp handle
SpO
2
pulse pitch tone, Battery status, External power LED
Interface
DC input connector: 18VDC, 2.8A
LAN digital output for transferring data,
Nurse call system connection
DC output: 5VDC, 1A Max
Battery
Rechargeable Li-ion battery
Continuous Battery Usage Time: 3 hours or more when fully
charged (6BL335-BIO-S, measured every 5 minutes NIBP with SpO2
and ECG)
Thermal Printer (option)
Speed: 25, 50mm/sec, Paper width: 58mm
Data Storage
168hours trends, 20 cases of 10 sec alarm waveform
Language
English, Korean, French, Bulgarian, Polish, German
Chinese, Portuguese, Hungarian, Czech, Romanian, Italian, Turkish,
Spanish, Russian, Greek, Japanese
ECG
Lead type
3-lead, 5-lead (option)
Lead Selection
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
ECG waveforms
3-lead: 1 channel
5-lead: 1/2/7 channels
Heart Rate Range
Horse, dog: 30 to 300 BPM,
Puppy, cat: 30 to 350 BPM
Heart Rate Accuracy
±
1bpm or
±
1%, whichever is greater
Sweep speed
6.25, 12.5, 25, 50 mm/sec
Filter
Diagnostic mode : 0.05Hz - 150Hz
Monitoring mode : 0.5
–
40 Hz
Surgical mode :0.5
–
25 Hz
S-T segment detection
range
-2.0 to 2.0 mV
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...