150
BM7Vet Pro
User’s Manual
A-4-12.
O2 TYPE
O2 Sensor type setup menu (default:
SERVOMEX)
Galvanic
Servomex
A-4-13.
IR O2 Delay
In seconds
A-4-14.
O2 OPTION
O2 function available setup menu
ON/OFF
A-5.
MODULE INFORMATION
Module Information Menu
A-5-1.
SENSOR SN
Serial Number of Module
PNXXXXXX
A-5-2.
H/W VERSION
Hardware Version of Module
1.2
A-5-3.
S/W VERSION
Software Version information
X
A-5-4.
COMM. PROTOCOL
Communication protocol
1.5
A-5-5.
O2 COMPENSATION
O2 Concentration Compensation
Vol%
A-5-6.
N2O COMPENSATION
N2O Concentration Compensation
Vol%
B-1.
ZEROING
Perform M-GAS zero point
C-1.
MODULE RESET
Reset M-GAS Module
Multi-gas Maintenance
The ISA sidestream gas analyzers are permanently factory calibrated and requires no
routine user calibration. For all ISA sensor fitted with an O
2
sensor, O
2
gas span
calibration instructions are described in these sections.
Calibration Gas
The gas span check and calibration requires the use of a suitable calibration gas mixture.
A cylinder containing such a mixture may be acquired from the vendor of your choice,
provided the following gas concentrations and accuracy. Depending on your application,
multi-gas calibration gas mixture shall be used
ISA Multi-gas
Gas
Gas Concentration
Summary of Contents for BM7VET
Page 69: ...69 BM7Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Page 156: ...156 BM7Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Page 219: ...219 BM7Vet Pro User s Manual SYSTEM LOWBATTERY...
Page 225: ...225 BM7Vet Pro User s Manual less than minus number percent plus or minus...