Append C Alarm Information
--150--
User Manual of Veterinary Monitor
minute
R on T
R on T detected within the last minute
Multiform PVCs
Ventricular premature of 2 or more
forms is detected in the last minute
HeartBeat Pause
Not detecting heart pacing within
preset cardiac arrest threshold time
Missed Beats
Unable to detect the heart pacing
within 1.75 times of the mean RR
period when the heart rate is <100, or
unable to detect the heart pacing in 1s
when the heart rate is >100.
Pacemaker NOT Capture
Asystole with pace-making pulse in
the last minute (Only applicable to
pacemaker-wearing animals)
The pacemaker has problem. Please check
the pacemaker.
Pacemaker Not Pace
No pace-making pulse detected within
a period that is 1.75 times the average
R-R intervals (Only applicable to
pacemaker-wearing animals)
Resp
Alarm Information
Triggering Condition
Treatment Measure
Resp Apnea(Resp)
No breathing signal within the preset
time of respiratory asphyxia
Check the animal state, electrode and lead
cable.
Resp Heatbeat Interrupt
The heart beat of the animal interferes
with the respiration.
CO
2
Alarm Information
Triggering Condition
Treatment Measure
Resp Apnea (CO
2
)
The animal has no breath, or the
breathing signal is too weak.
Check the animal‘s state, accessories and
airway connection.
Technical Alarm Information
Communication Module
Alarm Information
Triggering Condition
Treatment Measure
xx Communicate Error
Module not connected to host, or
initialization failed, or error with
module configuration
Restart the equipment. If the error remains,
please contact the manufacturer for repair.
xx
Communication
Stopped
Problem with the communication
between module and host
Restart the equipment. If the error remains,
please contact the manufacturer for repair.
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......