Chapter 15 Troubleshooting and Solutions
User Manual of Veterinary Monitor --129--
15.9 The Sound of Sidestream CO
2
Pump Becomes High (Optional)
Ensure the air tube is free of any foreign particles such as the water droplets, sputum or blood clots. Check if the
color of the filter wool inside the water trap is dark (brown or black). If yes, change the water trap. If the problem
remains, please contact the manufacturer.
15.10 Body Temperature without Numerical Value or Inaccurate
Readings
No value
First check whether the probe is inserted properly and then check whether the probe has physical fracture and
contact the manufacturer.
Inaccurate readings
First check whether the metal part of the probe sensor is in close contact with the tested part and then verify
that whether the measurement time is more than four minutes, ensure that the animal or the animal‘s position
being tested is essentially stationary; if fever cramps or convulsions cause the sensor loose, or axilla and
other parts that have dense body hair cause slow heat conduction or error, the hair should be shaven or select
other suitable positions for measurement.
Attention
If the machine has problems when you are using our monitor, you may check as described above. If
the problem remains, please contact the local dealer or call us directly.
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......