Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor --47--
PR: Pulse Rate as HR/PR alarm source.
Auto: when ECG measure is started and effective HR are obtained, the monitor will use ECG-derived heart
rate as its alarm source. In case heart rate is not available, e.g. when leads are not connected and there has
been one pulse source that is useable, the monitor will automatically use the pulse rate derived from the
current measure as the pulse source that is used as its alarm source. The monitor will then automatically
reuse the heart rate as its alarm source if the heart rate is obtainable.
Alarm Setup
Select [Alarm Setup>>], and [PAR.Alarm] will automatically pop up: HR/PR: which enable setting of on/off
of alarm switches, high and low alarm limits, alarm level, and whether or not to open the alarm record.
Other Setup
Select [ECG Setup]-[Others>>]:
QRS Volume:
0-10.0 and 10 indicate mute and the maximum volume, respectively.
Notch Filter:
when the Filter Mode is ‗Monitor‘ or ‗Surgery‘, the notch filter is defaulted open (50Hz or
60Hz); when the Filter Mode is ' Diagnose ', the notch filter can be opened or closed to meet different
demands.
Lead Set:
3-Lead and 5-Lead.
Screen:
select ECG working interface, and different options are available for different [Lead Set] settings:
1)
3-Lead: Normal.
2)
5-Lead: Normal, 7 Lead Half and 7 Lead Full.
Paced:
please refer to Check on Pacemaking Status for more details.
Save Curve:
I, II, III, aVR, aVL, aVF and V, namely, the ECG lead which allows selection of Long ECG
review storage.
ST Use:
ST Point, J+60 and J+80, and ST use mode can also be set.
Smart Lead Off:
in case that Smart Lead Off is selected, when the lead that is set in ECG or being displayed
on the interface is disconnected but there are other leads still useable, the system will automatically select the
useable leads and re-compute the heart rate. Still, when it is reconnected, the now connected lead will be
automatically restored to its original state.
6.1.8 ST Analysis
The ST analysis function is not suitable for animals < 10 kg.
ST analysis can measure ST elevation or reduction on a given lead.
Unit of ST measured value: mV or mm.
Meaning of ST measured value: a positive number indicates an elevation while a negative one indicates
a reduction.
ST measuring range: (-2.0mV)-(+2.0mV).
Warning
ST data should not be used as a judgment standard for doctors and its clinical significance is
decided by the doctors.
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......