Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor --63--
PR Source Setup
Select [PR Source]: SpO
2
, IBP1, IBP2 and Auto. The system automatically switches the [PR Source] to
[Auto] when the current PR source is not available.
Alarm Source Setup
As referred to Alarm Source Setup in section 6.1.6.
Pulse Volume Setup
Select [Pulse Volume]: 0-10.
Sensitivity Setup
The SpO
2
values displayed on the monitor are results achieved by averaging data sampled within a given
period of time. A shorter averaged time represents a higher response speed of the monitor yet lower
measuring accuracy when the animal‘s SpO
2
values are changing. Correspondingly, a longer averaged time
represents lower response time of the monitor yet higher measuring accuracy when the animal‘s SpO
2
values
are changing. A relatively shorter averaged time is conducive to analysis of conditions of critical animals that
are being monitored.
SpO
2
module: select [Sensitivity]: High, Mid and Low, and corresponding averaged time: 7s, 9s and 11s.
Pump Show Setup
Select [Pump Show]: On or Off.
[On]: pump Show is displayed in SpO
2
parameter region.
[Off]: pump Show is not displayed in SpO
2
Parameter region.
Wave Fill Setup
Select [Wave Fill]: On or Off.
[On]: Pleth waveforms on screen are displayed being filled.
[Off]: Pleth waveforms on screen are displayed not being filled.
Alarm Setup
Select [Alarm Setup>>], and SpO
2
-matched [PAR.Alarm] will automatically pop up, which enable Setup of
on/off of alarm, higher and lower alarm limits, alarm levels, and whether or not to open the alarm record.
6.4.5 Influencing Factors of Measure
Animal‘s body movement
Electromagnetic influence, e.g. by NMR equipment
Electro surgical equipment
Existence of some dyes
Outside optical radiation
Improper placement of blood oxygen sensor or use of incorrect blood oxygen sensor
Animal‘ s shock, anemia and low temperature
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......