Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor --57--
Figure 6.10 Resp Setup
Apnea Delay Setup
Select [Apnea Delay]: 10s, 15s, 20s, 25s, 30s, 35s and 40s.Its default is 10s.When apnea delay is set, the
monitor will alarm in case the apnea period exceeds the preset values.
Gain Setup
Select [Gain]:
×
1/4,
×
1/2,
×
1,
×
2,
×
4 and Auto. It allows setup of waveform gain positions to meet
different clinical requirements.
Sweep Setup
Select [Sweep]: 6.25 mm/s, 12.5 mm/s, and 25.0 mm/s. It allows selecting desired waveform speeds. A larger
speed means faster scanning and wider waveforms.
Resp Lead Setup
Select [Resp Lead]: I and II. It allows selecting the desired lead to meet different clinical demands for
optimal respiratory waves.
Attention
In order to obtain optimal respiratory waves, RA and LA electrodes should be kept level when I-lead
is selected for resp measure, and RA and LL electrodes should be diagonally kept when II-lead is
selected.
Detect. Mode Setup
Select [Detect. Mode]: Auto and Manual. Respiratory wave calculation mode can be set.
Auto Mode
In this mode, the monitor will automatically adjust the detection level according to waveform height and
whether cardiac artifact is existent, but the detection level dotted line is not displayed on the respiratory
waveforms.
[Auto] mode is recommended when respiratory rate is not close to the heart rate or animals are actively
respiration.
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......