Chapter 14 Maintenance
User Manual of Veterinary Monitor --123--
Chapter 14 Maintenance
14.1 Safety Information
Warning
The removal or repair of the monitor can only be done by the well-trained professional technicians.
If you find any problems, please contact us or repair technician.
14.2 NIBP Accuracy Test
Refer to 6.5.9 for details.
14.3 NIBP Overpressure Test
Select [Main Menu]-[Maintenance]-[NIBP Overpressure Test] The characters on the key are changed to [Stop
NIBP Overpressure Test]. The ‗Overcharge Testing…‘ is displayed on NIBP parameter window. Select [Stop
NIBP Overpressure Test] or press [NIBP] key on the panel to manually stop the NIBP Overpressure Test.
Warning
NIBP Overpressure Protection Test: it should not exceed 300 mmHg (39.9 kPa) for the animals > 20
kg, not exceed 240 mmHg (31.9 kPa) for the animals from 10~20 kg and not exceed 150 mmHg
(19.9 kPa) for the animals > 20 kg.
14.4 NIBP Leakage Test
Refer to 6.5.8 for details.