Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor --51--
PVC
Single PVC is detected in normal heartbeat.
Alarm)
Tachycardia
The average heart rate is greater than the limit of the
tachycardia.
Bradycardia
The average heart rate is lower than the limit of
bradycardia.
VR(ventricular rhythm)
Adjoining dominant rhythm of ventricular heart
beats exceeds idioventricular rhythm threshold
numbers and heart rate is lower than ventricular
tachycardia (VT).
V-Bigeminy
Rhythm N, V, N and V.
V-Trigeminy
Rhythm N, N, V, N, N, V.
Irr.Rhythm
Continuous irregular rhythm.
PVCs/min
PVCs/min exceeds preset higher limit.
Run PVCs
>
2
More than 2 continuous PVCs in the last minute.
Couplet
Paired PVCs detected in the last minute.
R on T
R-wave on T-wave detected in the last minute.
Multiform
Ventricular premature of 2 or more forms is detected
in the last minute.
HeartBeat Pause
Heart beat not detected when preset asystole
threshold time has passed.
Missed Beats
In case the heart rate is below 100, heart beat is not
detected within a period that is 1.75 times the
average RR intervals or in case the heart rate is
above 100, heart beat is not detected in 1 second.
PNC
(Pacemaker
Not
Capture)
Asystole with pace-making pulse in the last minute
(Only applicable to pacemaker-wearing animals).
PNP
(Pacemaker
Not
Pace)
No pace-making pulse detected within a period that
is 1.75 times the average R-R intervals (Only
applicable to pacemaker-wearing animals).
Arrhythmia Alarm Setup
Select [Alarm Setup]-[Arrh.Analysis], or select ECG parameter region -[Arrh.Analysis>>], and then set
alarm setting for each sort of arrhythmia in the pop-up menu. Select alarm switch for each option to start or
stop corresponding arrhythmia analysis function; you can also switch on/off the arrhythmia analysis alarm by
selecting three functional buttons below the menu: [Lethal Only], [All Off], and [All On].
Attention
Fatal arrhythmia defined in this series of monitors include: Asystole, and VFib/VTac (ventricular
fibrillation/ ventricular tachycardia).
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......