Chapter 7 Alarm
User Manual of Veterinary Monitor --95--
Attention
When the system is in the
‘Alarm Silence’ state, any newly triggered alarms will release the ‘Alarm
Silence
’ state, but only the new one has normal audible and lighting alarms, leaving the silenced
alarm still being silenced.
Disconnect the probe module technology caused alarm cannot be turned off, the alarm will always
exist until the probe is re-connected, the module reload successfully, and then alarm disappears.
7.9 Alarm Detection and Counter Measures
The monitor will perform alarm self-check when it starts. At that moment, the red and yellow alarm lights are
turned on in turn and are turned off simultaneously when the system has a ‗thudding‘ sound. This means that the
audible, lighting alarm indicators work normally.
When the monitor gives alarms:
1.
Check the animal‘s actual clinical condition.
2.
Confirm the type and parameter of the current alarm.
3.
Recognize the alarm cause.
4.
Deal with the alarm cause.
5.
Check whether the alarm disappears.
Please refer to the
Alarm Information
listed in
Appendix C
for detailed counter-measures of each alarm.
7.10 Other Bed Alarm
7.10.1 Other Bed Alarm Auto Prompting
Set the function of automatic prompting for other-bed alarms:
1.
Select the shortcut key [Screen] on the main screen-[Screen Config]-[Interface Type]-[View Other Bed].
2.
Select the button [Setup] in the View Other Bed window and set the [Auto Alarm]-[On] in the pop-up menu.
In case the other bed alarm automatic prompting function is set to On, the monitor also provides prompting
information in the Prompting Message Region as shown in Figure 7.4 when another-bed monitor alarms but
its observation interface is not started in another-bed cluster established by the monitor.
Figure 7.4 Other Bed Alarm Prompting Information
7.10.2 Other Bed Alarm Silence
You can perform remote alarm silence control for the currently observed other-bed monitor in the other-bed
observation interface.
In case the other-bed mute function is set [On], press the button [SILENCE] in the other bed observation interface
and then the current alarms of the currently observed other-bed monitor can be silenced when it alarms.
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......