Append B Product Specifications
--138--
User Manual of Veterinary Monitor
electrosurgical knife is not activated.
CMRR
Diagnose Mode:
≥
89 dB
Surgery & Monitor Mode:
≥
100 dB
Time Constant
Monitor Mode:
≥
0.3 s
Diagnose Mode:
≥
3.2 s
Frequency Response
Surgery Mode: 1 Hz-15 Hz; Monitor Mode: 0.5 Hz-40 Hz; Diagnose Mode:
0.05 Hz-150 Hz.
ECG
Parameter
Frequency
Characteristics
Surgery Mode: Meet (
+
0.4 dB
~
(-3.0 dB)) requirements at 15 Hz.
Monitor Mode: Meet (
+
0.4 dB
~
(-3.0 dB)) requirements at 0.5 Hz
~
40
Hz.
Diagnose Mode: Meet (
+
0.4 dB
~
(-1.0 dB)) requirements at 0.05 Hz
~
60 Hz.
Meet (
+
0.4 dB
~
(-3.0 dB)) requirements at 61 Hz
~
150 Hz.
Notch
Monitor & Surgery Mode: notch filter automatically activated at 50 Hz/60
Hz
Diagnose Mode: Notch filter manually activated or deactivated at 50 Hz/60
Hz
Range of Electrode Polarized
Voltage
±300 mV d.c.
Lead Fall Testing Current
Measuring Electrode:
<
0.1 μA
Drive Electrode
<
1 μA
Pacemaker Pulse
Pacemaker
Pulse
Display
Capacity
Pace-making mark can be displayed for the following pacemaker pulses:
Pulse Amplitude: ±2 mV
~
±100 mV
Pulse Width: 0.1 ms
~
2 ms
Pulse Rise Time: 10 μs
~
100 μs
Pacemaker pulse should be no overshoot
Pacemaker Pulse Suppression
Capacity
The monitor can inhibit the pacemaker pulse that conforms to the following
conditions:
Pulse Amplitude: ±2 mV
~
±100 mV
Pulse Width: 0.1 ms
~
2 ms
Pulse Rise Time: 10μs
~
100 μs
Pacemaker pulse should be no overshoot
Alarm Limit Specifications
Range
Upper Limit of ECG Heart Rate
Alarm upper limit for
>
20 kg: (Lower limit+2) bpm
~
300 bpm
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......