Appendix E EMC- Guidance and Manufacture‘s Declaration
User Manual of Veterinary Monitor --165--
Appendix E EMC- Guidance and Manufacture
’s Declaration
E.1 Guidance and manufacture
’s declaration-electromagnetic emissions
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The Veterinary Monitor is intended for use in the electromagnetic environment specified below. The customer
or the user of Veterinary Monitor should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
EN 55011
Group 1
The Veterinary Monitor uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emissions
EN 55011
Class A
The Veterinary Monitor is suitable for use in all establishments,
not including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
EN 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
EN 61000-3-3
Complies
E.2 Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Veterinary Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of the Veterinary Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrostatic
transient / burst
EN 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power
supply lines
±
1
kV
for
input/output lines
Mains power quality should be that of a
typical
commercial
or
hospital
environment.
Surge
± 1 kV differential
mode
±
1
kV
differential mode
Mains power quality should be that of a
typical
commercial
or
hospital
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......