Append B Product Specifications
User Manual of Veterinary Monitor --141--
Tolerance
higher.
The tolerance is not defined for other ranges.
Asphyxia Alarm Tolerance
Within 10 s
~
40 s (Increase/decrease by 5s for each rotation of the knob), the
asphyxia alarm tolerance is ±5 s.
Alarm Limit Specifications
Range
RR Upper Limit
Alarm upper limit for
>
20 kg: (Lower limit+2) rpm
~
100 rpm
Alarm upper limit for 10
~
20 kg: (Lower limit+2) rpm
~
100 rpm
Alarm upper limit for
<
10 kg: (Lower limit+2) rpm
~
100 rpm
RR Lower Limit
Alarm lower limit for
>
20 kg: 0 rpm
~
(Upper limit-2) rpm
Alarm lower limit for 10
~
20 kg: 0 rpm
~
(Upper limit-2) rpm
Alarm lower limit for
<
10 kg: 0 rpm
~
(Upper limit-2) rpm
B.6.3 SpO
2
Monitoring
Alarm Limit Specifications
Range
SpO
2
Upper Limit
(Lower limit+1)%
~
100%
SpO
2
Lower Limit
80%
~
(Upper limit-1)%
Accuracy Tolerance
±1% of the setting
Sensing element
Optical power <15 mW
Red light wavelength: 658 nm~664 nm, infrared light: 897 nm~915 nm
Information on the wavelength range is particularly useful for clinicians (e.g.
in optical dynamic therapy)
Monitoring Parameters
SpO
2
and Pulse Rate (PR)
Range
0%
~
100%
Resolution
1%
Data update peiriod
1 s
Accuracy
Within 70%
~
100%, the measuring tolerance is ±2%.
Within 0%
~
69%, the measuring tolerance is not defined.
B.6.4 PR Specifications
Alarm Limit Specifications
Range
PR Upper Limit
Alarm upper limit for
>
20 kg: (Lower limit+2) bpm
~
250 bpm
Alarm upper limit for
<
10
~
20 kg: (Lower limit+2) bpm
~
250 bpm
Alarm upper limit for
<
10 kg: (Lower limit+2) bpm
~
250 bpm
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......