Append B Product Specifications
--148--
User Manual of Veterinary Monitor
Accuracy
When
<
5.0%: ±0.3% (±2.0 mmHg) (0.27 kPa)
When
≥
5.0%:
<
6% of the reading
Range of Breathing Rate
3 rpm
~
150 rpm
Accuracy of Breathing Rate
1% or ±1 rpm, whichever is higher.
Asphyxia Alarm Delay
30 s, 35 s, 40 s, 45 s, 50 s, 55 s, 60 s
Automatic Pressure Compensation
yes
Alarm Limit Specifications
Range
EtCO
2
Upper Limit
(Lower Limit +2)mmHg
~
99 mmHg
EtCO
2
Lower Limit
0 mmHg
~
(Upper Limit -2)mmHg
FiCO
2
Upper Limit
0 mmHg
~
99 mmHg
awRR Upper Limit
(Lower limit+2) rpm
~
100 rpm
awRR Lower Limit
0 rpm
~
(Upper limit-2) rpm
B.6.9 Recorder Specifications (Optional)
Recorder
To record the animal information, the hospital information, waveform,
parameters and others displayed in the screen
Method
Thermal array recorder
Printing Paper
Thermal paper
Print Resolution
8 dots/mm on Y-Axis
Delay Characteristics
≤
0.5 mm
Amplitude-frequency Characteristics
Monitor Mode: 0.5 Hz
~
40 Hz; Diagnose Mode: 0.05 Hz
~
150 Hz.
Time Constant
≥
0.3 s
Summary of Contents for PM-900
Page 2: ......
Page 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Page 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Page 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Page 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Page 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Page 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Page 181: ......