S3 Cell Sorter Instruction Manual
90 |
If sonication of the tip is not performed:
10. Using a 1 ml syringe filled with ethanol, press against the top and eject ethanol through the
nozzle tip, rotating between the 100 µm tip and the larger opposite end to free clog.
11. Watch the resulting stream. The fluidic stream should be coming out straight from the tip.
12. Repeat if necessary until the stream is coming out straight.
13. Proceed to step 14.
14. Place the O-ring at the point on the nozzle where the bottom of the black collar will be.
15. Place the nozzle tip back onto the nozzle assembly by pushing it up using the sides of the
tip. Avoid touching the end.
16. Secure the black retaining collar until finger tight.
17. Remove the lint-free cloth from nozzle chamber.
18. Press the stage back down and tighten the thumb screw.
19. Click the blue forward arrow.
Fig. 101. Nozzle swap wizard window.
Fig. 100. Nozzle swap wizard window.
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......