S3 Cell Sorter Instruction Manual
| 21
Table 20. Instrument controls and their functions.
Button
Name
Function
Start Sample
Starts sample acquisition. If a data file is loaded and it has been saved,
Acquisition
the data will clear and acquisition will start
Stop Sample
Stops acquisition and sample flow. Regions, gates, and annotations may
Acquisition
be modified before saving the FCS file
Start Sorting
Starts sorting. If a data file is loaded and it has been saved, the data will
clear and sorting will start
Stop Sorting
Stops sorting and sample flow. Regions, gates, and annotations may be
modified before saving the FCS file
Clear Acquired Data Clears acquired data. All plots and statistics will be refreshed.
Pause Sample
Pauses sample acquisition and sample flow. The data at this point may be
appended (by the Resume function) or saved as is. Saving after pausing will
disable the Resume function. To remove sample tube, unlock the touchscreen
locking system located above the loading stage after pausing acquisition
Resume Sample
Resumes acquisition when a sample has been paused. Resuming will
append the paused FCS file and the time paused can be seen on a time
histogram. If you have removed the sample tube it must be returned and
relocked before acquisition can be resumed
Eject Sample Tube
Depressurizes and unlocks the sample. The loading stage can be moved
into wash position to access sample. The touchscreen locking system can
also be used to unlock the sample
Adjust Sample
Starts or stops agitation of the sample. There are three levels of agitation:
Agitation
low, high, and off. Click this button once to set to low, twice to set to high,
and three times to turn off agitation
Auto Save
When checked, automatically prompted to save a data file after
acquisition stops.
Target Event Rate
Allows selection of target event rate for sample acquisition or use of preset
sample rates low/med/high based on sample concentration
Limit
When checked, stops data acquisition after the number of events entered
has been reached
Gate Limit
When checked, data acquisition stops after the number of events entered
has been reached within the selected gate
Cycle Mode
When checked, automatically refreshes acquired data at the entered time
Sample Chamber
Turns the sample chamber light on and off
Light
Enable/Disable Sort Illuminates entire sort collection chamber for easy viewing of deflection
Chamber Light
plates, sort stream, and collection vessels
Enable/Disable
Enables or disables temperature control of the sample station and sort
Temperature Control collection vessel
Biosafety
Indicates the status of the biosafety system and gives information on filter life.
System Status
If the biosafety system loses connection, cannot connect, or if the filter life is
at 0% an error will occur and not allow the user to sort or acquire.
continues
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......