S3 Cell Sorter Instruction Manual
| 33
Getting Started
4
After system installation by a Bio-Rad certified service engineer, the S3 system should not be
turned off via the main power switch. The system is safe in standby mode after performing a
software shutdown of the system. This will help facilitate quick or automated startup.
If the system is shut off via the main power switch, always power the instrument on before
starting the ProSort
™
software. The internal computer system of the instrument must be running
in order to communicate with the software.
IMPORTANT!
Powering the system on or off incorrectly can cause the instrument to
not run properly.
4.1 Checking Bulk Fluidics
Prior to starting the ProSort software, open the bulk fluidics door and check the volume
of each container. Ensure that the waste container is empty, and the sheath fluid and DI
containers are full.
Note:
To change the bulk fluidics after starting the software or during a run, refer to Section 5.3,
Checking or Swapping Fluidics.
To refill or change a container (Figure 29):
1. Detach the quick disconnect from the cap assembly by pushing in the metal button and
lifting up the quick disconnect until it engages the magnetic holder.
2. Slightly lift and pull out the container.
3. Remove the cap assembly from the container.
4. Refill or change the container.
5. Carefully place the cap assembly onto the container and tighten the cap.
6. Place the container into position.
7. Attach the quick disconnect to the cap assembly. An audible click should be heard.
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......