S3 Cell Sorter Instruction Manual
100 |
Table 29. Troubleshooting table (continued).
Error
Possible Causes
Troubleshooting Steps
Drastic light change
Background movement in
Redo QC and avoid changing light
front of green sliding door
conditions in front of green sliding door
during QC process
during the procedure
Sheath error
The internal sheath
The connections to the DI water and 8x
reservoir cannot maintain
sheath fluid containers should be checked to
proper levels
make sure they are connected, that they
are not empty, and that fluid can flow. It is
possible that their filters are clogged and
need to be replaced
Waste error
Waste quick disconnect
The connections to the waste container
not connected properly to
should be checked to make sure they are
waste container
connected and that waste can flow. Once
corrected, it will take time for the system to
Internal waste reservoir filled
clear the internal reservoir. You can observe
that pumping is happening by watching for
splashing in the waste container
Nozzle stage cannot return to
Laser intersection cover has
Adjust laser intersection cover so it is seated
proper position after nozzle
been moved during tip swap
correctly. Once in the proper place the
tip swap
nozzle stage can be lowered into position
Cannot acquire sample.
Room too cold or faulty
Warm room to >63°F. If this does not help,
Hovering on acquisition button
bubble detector
contact service support
displays “bubble detected”
Arcing
The high voltage deflection
Stream spray or droplets are on the
plate monitor has measured
deflection plates. Carefully clean and dry
the plate voltage deviating
the plates
from normal
No data appear on the work-
Nozzle interlock has been
Open nozzle door all the way then close to
space after changing nozzle tip
activated
reset the laser interlock
No events during acquisition
Nozzle door open
Close the nozzle door
Laser shutter closed
Cycle the nozzle door open then closed,
especially if cover was removed
Laser not on
Turn on the laser in the user interface
Restart instrument and computer to correct
bad communication
Threshold too high
Lower threshold
Triggering on noise
Change the trigger parameter to FSC or SSC
Stream not centered in
Use the electronic micromotors (Agillis)
laser beam
positioner while acquiring to maximize event
signal — access within the system log file
No beads in sample
Check sample tube, vortex
continues
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......