S3 Cell Sorter Instruction Manual
74 |
7.4 Clean System
In between samples or users, cleaning can be done at low or high pressure. High pressure
cleaning will cause the system to stop maintaining droplets and a QC procedure must be run
before sorting. If cleaning between samples, the low pressure option is recommended. The
fluidics lines will be flushed thoroughly with cleaning solution, similarly to the cleaning protocol
used during the shutdown or logout procedure.
To clean the system:
1. Fill a 5 ml tube with a minimum of 3 ml of cleaning solution.
2. Load tube into the sample station and move in to run position.
3. Click the
Set Up and Maintenance
tab.
4. Click
Clean System
.
The cleaning protocol will automatically begin. A prompt will appear to warn the user that the
clean function will begin and that the contents within the sample tube will be used rapidly.
Fig. 78. Cleaning protocol warning box.
Fig. 79. Clean System modes.
Fig. 80. Warning for high pressure cleaning protocol.
5. Select the cleaning mode to run (Figure 79).
6. Click
Yes
to proceed with mode (Figure 80).
7. Move the loading stage into wash position (Figure 81).
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......