S3 Cell Sorter Instruction Manual
54 |
5.1.7 Auto Save
An auto save option is available as a check box on the instrument control panel. If this box is
checked (Figure 58), a prompt will appear to name and save the FCS file after an acquisition or
sort is completed.
Fig. 58. Checked auto save box.
Fig. 57. PMT control panel with name
changed on the FL1 parameter.
5.1.5 Threshold
The threshold level is set using the trigger parameter with a range of 0.1–100% (Figure 56).
The purpose of the threshold is to set an intensity level to eliminate any background noise and
debris while still allowing real events to be detected and analyzed.
5.1.6 PMT Voltages
Each parameter is collected by a photo multiplier tube (PMT). Independent voltage control is
available for each PMT. These voltages can be adjusted to bring the populations of interest
on the histogram or density plot scale (Figure 57). Custom names for each parameter can
be entered by clicking on the parameter of interest on the PMT control panel. Changing this
name will update the names on all plot axes, in the filter configuration window, and on the
compensation matrix. Refer to Section 3.3.8, Global Preferences, for information on how to set
default parameter names.
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......