| 9
S3 Cell Sorter Instruction Manual
| 9
The sort positions for each collection vessel are numbered and will correlate to the position
numbering in the software when the sort logic and limits are set. When placing tubes in the
sort chamber for sorting, it is recommended to add media or buffer to the tube to help prevent
sorted cells from drying out and to cushion their collection. The minimum recommended
volume is 0.5 ml of media or buffer for each 5 ml tube. Quick-attaching adaptors are available
for the 8-well strips or microscope slides. Adaptors are part of the accessory kit and they can
be kept inside the fluidics door for storage.
CAUTION! Biohazard!
When running samples and sorting, hazardous aerosols may be
created depending on the sample type. To prevent hazardous aerosols from spreading,
keep the green containment door closed as often as possible. Consult with your local
safety officer or review local, state, and federal regulations to ensure proper handling and
disposal of biohazardous substances including samples, sorted fractions, and waste.
CAUTION! Shock hazard!
The sort deflection plates in the sort chamber are
charged while sorting. An interlock on the sort chamber door will disable the plates
when opened. Only qualified personnel should override this interlock.
2.3 Optics
The S3 optics include the laser(s), mirrors, filters, and lenses, which guide the laser light to the
stream of sample and collect scattered and emitted light for detection.
CAUTION! Laser hazard!
Laser radiation can be hazardous. Please do not override
optical interlocks or remove light shields, as they are in place for your safety. Only
trained personnel should access the exposed laser beam.
2.3.1 Laser
The S3 system is installed with a 488 nm, 100 mW laser for cell and particle interrogation that
passes through the system in the sample core. The laser power and shutter can be controlled
through the software interface. An optional 561 nm or 640 nm 100 mW laser are available as a
second excitation source.
Fig. 9. Sort collection chamber and deflection plates.
Deflection plates
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......