S3 Cell Sorter Instruction Manual
78 |
IMPORTANT!
The S3 biosafety system must remain on at all times when the S3 cell
sorter is inside the system. The cell sorter should be left with the main system power
on and will enter into idle mode when not in use.
Should the biosafety system be turned off when the S3 cell sorter power is on inside,
there could be risk of overheating and it may become a fire hazard.
If any of the following occur, the S3 cell sorter will stop sample acquisition or sort and alert
the user:
n
Airflow through HEPA filter drops below required limit
n
Filter life reaches 0%
n
ProSort software loses communication with the enclosure
Until resolved, the software will not allow running samples or sorting.
When performing the swap tip procedure, the front panel may be lifted and attached via magnet
to the upper filter housing to allow nozzle access. Note that in this state, the S3 biosafety system’s
airflow is reduced and the system is not considered to be in Class I containment.
IMPORTANT!
If the airflow through the HEPA filter falls below required limits
for containment, the software will alert the user and stop the sort or
acquisition immediately.
7.10 Biosafety System
The S3
™
biosafety system is a Class I aerosol containment hood designed to work with the
S3 cell sorter, which works in conjunction with ProSort
™
software, providing users and the
environment protection from aerosols created during the cell sorting process.
The biosafety system is fully integrated with and monitored by the S3 cell sorter’s ProSort
software, so users have real-time information about HEPA filter life and the temperature of the
system inside. To maintain containment, the fan speed is regulated based on the cell sorter
operation mode and the airflow rate through the HEPA filter.
The S3 biosafety system functionally meets airflow requirements from the National Institutes of
Health (NIH) standards for Class I biosafety and the International Society for Advancement of
Cytometry (ISAC) guidelines.
For more information please see the S3 Biosafety System Class I manual.
Fig. 87. Biosafety System Class I status window.
Summary of Contents for 145-1001
Page 1: ...S3 Cell Sorter Instruction Manual Catalog 145 1001 145 1002...
Page 2: ......
Page 8: ...S3 Cell Sorter Instruction Manual vi vi...
Page 10: ...S3 Cell Sorter Instruction Manual viii...
Page 22: ...12 S3 Cell Sorter Instruction Manual 12...
Page 42: ...S3 Cell Sorter Instruction Manual 32...
Page 60: ...50 S3 Cell Sorter Instruction Manual 50...
Page 74: ...S3 Cell Sorter Instruction Manual 64...
Page 82: ...72 S3 Cell Sorter Instruction Manual 72...
Page 96: ...86 S3 Cell Sorter Instruction Manual 86...
Page 112: ...102 S3 Cell Sorter Instruction Manual 102...
Page 114: ...S3 Cell Sorter Instruction Manual 104...
Page 116: ...106 S3 Cell Sorter Instruction Manual 106...
Page 118: ......
Page 119: ......